A spore test (biological indicator) is the only check that proves your sterilizer actually killed the hardest thing to kill. Chemical indicators and the cycle printout show the conditions were reached; the spore test shows they worked.
The schedule
The CDC recommends biological monitoring at least weekly, and with every load that contains an implantable device. Many offices test more often, and some states require a specific interval, so your state's rule wins if it's stricter.
Run the test strip or vial inside a normal, full load, placed where the sterilizer or indicator manufacturer says to put it (often the center, or the spot hardest for steam to reach). Run a control from the same lot that was not sterilized alongside it so you know the spores were viable to begin with.
When it fails
A positive result means spores survived. Don't panic and don't ignore it. Take the sterilizer out of service, review the cycle (loading, packaging, time, temperature, pressure), and retest. If the repeat test passes and the mechanical and chemical indicators were normal, you can return the unit to service. If it fails again, the sterilizer stays down until it's serviced and passes three consecutive spore tests, run the way CDC and the manufacturer specify (CDC's current guidance says fully loaded chamber cycles). One exception to the don't-panic rule: anything implantable from a load with a positive result is recalled and quarantined until it's shown to be sterile, even if the retest passes.
If the repeat test is also positive, or the failure is confirmed, everything processed since the last negative test is treated as non-sterile. Recall and reprocess what you can, and document everything: the failure, the actions, and the retests. The log is what proves you handled it correctly.